Fully Auto Clinical Chemistry Analyser
DC 400 |
Fully Automatic Clinical Chemistry Analyzer |
| Highly-Efficient Sample Processing, Constant Speed 400 T/H |
| Constant test speed of 400 T/H, optional ISE |
| 90 Sample Positions, 80 Reagent Positions |
| 90 Quartz Cuvettes |
| Dry Heating Thermostatic System - Rapid Heating, Maintenance Free |
| Directly Heating System Enables Faster Heating and Decreases Pre- Heating Time |
| Thernostatic System by Dry Heating Keeps Reaction Solution in Cuvettes at (37.0+0.2)'C. |
| Accurate Sampling System |
| High-Precision Injector |
| Internally and Externally Polished Sampling Probles |
| Liquid Volume Tracking and Leval Detection |
| Minimum Reaction Volume: 100ul |
| Minimum Sampling Volumes: 1.5ul |
| CV<2% |
| Intelligent Online Loading System |
| Reagents can be Loaded On-board During Test |
| Peltier System Provides a Longer Maintenance Period to Reduce Repair Probablity. |
| 8-Step Warm Liquid Washing - More thorough cleaning |
| Whole Probe Stero Cleaning Technology with Cross-Contamination Rate<0.05% |
| Pure Water Washing in Step 1, Cleaning Solution Washing in Step 2-3, Pure Water Washing in Step 4-6, Than Wiping Module Washing in Step 7-8 |
| Anti-overflow Function Prevents Liquid Spillage |
| Advanced Operation System |
| lntuitive user-friendly interface, easy to learn and use |
| Auto sample dilution |
| Auto stopping and warning when reagents or samples run out |
| Calibration with linear or non-linear curves. Multi-point calibration |
| Automatically detecting cuvettes status and screening flawed cuvettes |
|
Test Principle |
Latex Immunoturbidimetry, Immunoturbidimetry, Colorimetry |
|
Analysis Method |
Rate, End-point, Fixed-time |
|
Test Items |
Up to 78 items at the same time |
|
Analysis Rate |
Constant Speed 400 T/H |
|
Sample Tray |
90 Sample Positions (including sample, calibrator and controls) |
|
Sample Volume |
1.50-50.0 µL (0.1 µL increment) |
|
Reagent Volume |
R1: 20.0-300.0 µL |
|
R2/R3/R4: 0 or 20.0-300.0 µL (1 µL increment) |
|
|
Reagent |
Up to 4 reagents per test item |
|
Reagent Tray |
Reagent Positions: outer: 40 sample positions |
|
Reaction Volume |
100.0-360.0 µL |
|
Reaction Temperature |
Constant temperature control (37.0±0.2)°C |
|
Cuvette |
Quartz glass: 5*5mm; 90 cuvettes |
|
Weight |
300 kg |
|
Dimension |
1160 mm (L) x 790 mm (W) x 1140 mm (H) |
|
Water Consumption |
No more than 20 L purified water per hour |